Dimensions:
Desktop variant: 450 x 200 x 440 (w x h x d)Rack variant: U=4, HP=84, D=400mm
Weight:
Approx. 15 kgProtection type:
IP 20 to IEC 529Power supply:
230V +4% / -10%, 50-60Hz or115V ± 10%, 50-60Hz
Power:
Approx. 100W / Slow Fuse 2A in Power switchHeat up time:
Approx. 15 min.Climate:
Storage: 0…50 °COperation: 5…40 °C
Measuring range:
NH3: 0…20.000,0 ppmMeasuring gas components:
NH3Measuring method:
TDLS – Tunable Diode Laser SpectrometryAdditional measurement:
Temperatur TaupunktThermocouple K, S
Influence values:
Accuracy: ± 2% of reading or, ± 2 ppm whichever is largerPrecision 2σ : ± 0.9 ppm
Linearity: included in the accuracy
Repeatability: included in the accuracy
Zero drift over 2 h period:: below the accuracy
Operating pressure: 800 … 1100 mbar
Gas inputs:
Measurement gas inputZero gas input
Reference gas input
Pressure
Maximal pressure: 100 mbar overpressure (ü)Working pressure : 0 do 50 mbar overpressure (ü)
Gas flow: Approximately 3 l/min
Gasausgänge
Gas output 1Gas output 2
Pressure: Pressureless